What’s Wrong With My Chromatography - Agilent Technologies
What's Wrong With My Chromatography
Rita Steed LC Columns Application Engineer January 13, 2015
Confidentiality Label January 13, 2015 1
Outline of Presentation
I. Common Problems, Symptoms, and Solutions
- Baseline Shifts/Noise - Peak Shapes - Pressure - Retention Shifts
II. Preventative Maintenance/Good Practices
- In-line Devices - Sample Preparation - Operational Limitations
Confidentiality Label January 13, 2015 2
Common Problems
Noisy Baselines
Possible Causes
? Dirty flow cell ? Lamp failing ? Pulses from pump (if periodic) ? Temperature effects on detector ? Air bubbles passing through detector
Time (min.)
Confidentiality Label January 13, 2015 3
Non-spec Lamp
Chromatographic Results with Second Source Lamp at 214 nm
Lamp from Agilent Technologies
Peak 1 Peak 2 Peak 3
S/N = 150 S/N = 400 S/N = 300
Lamp from Second Source
Peak 1 Peak 2 Peak 3
S/N = 15 S/N = 50 S/N = 50
Drifting Baseline
Possible Causes
? Gradient elution
o Mobile phase component
? Temperature unstable (RID) ? Contamination in mobile phase ? Mobile phase not in equilibrium with
column ? Contaminant bleed in system
o Hardware chemical compatibility
Confidentiality Label January 13, 2015 5
Effect of TFA on Baseline
A: 0.1% TFA in H2O B: 0.1% TFA in ACN
Temperature: 35?C
Sample: 1. Phenacetin 2. Tolmetin 3. Ketoprofen 4. Fenoprofen 5. Ibuprofen 6. Phenylbutazone 7. Mefenamic acid 8. Flufenamic acid
Eclipse XDB-C8 4.6 x 150 mm, 5 m
Gradient: 5 - 100% B in 30 3 min
Flow Rate: 2.0 mL/min
6 8 7
Wavelength ?215nm
?254nm less impact
2
5
1
4
Adjust TFA concentration in Solvent B to level baseline
0
5
10
15
20
Time (min)
Know the UV Cutoff of your mobile phase components
January 13, 2015Month ##, 200X A.WennerGroup/Presentation Title
Agilent Restricted Page 6
Peak Shape
Ghost Peaks ? Appear When No Sample is Injected
60
20% - 100% MeOH
15
No Sample Injected
30
15
0
3
7
Problem ? Dirty mobile phase ? Sample carryover
? May imply poor recovery
? Pe15ak from17an early run (isocratic)
Confidentiality Label January 13, 2015 7
Solvent Contamination
___ Solvent Source 1 ___ Solvent Source 2 ___ Solvent Source 3 ___ Solvent Source 4, Lot 1 ___ Solvent Source 4, Lot 2
Confidentiality Label January 13, 2015 8
................
................
In order to avoid copyright disputes, this page is only a partial summary.
To fulfill the demand for quickly locating and searching documents.
It is intelligent file search solution for home and business.
Related download
- lab 4 column chromatography grossmont college
- how to run column chromatography
- ion exchange chromatography columns and resins
- chromatographic theory umass
- o fe ferrocene fluorenone
- hplc high pressure liquid chromatography introduction
- column chromatography purification of acetylferrocene
- th semester theory dse4t research methodology and design tools
- chromatography national council of educational research and training
- column chromatography
Related searches
- what s wrong with an annuity
- what s wrong with pete davidson
- what is wrong with my cat
- what s wrong with my car
- what s wrong with education today
- what s wrong with the internet today
- what s wrong with public education
- what s wrong with america today
- what s wrong with american education
- what s wrong with schools today
- what s wrong with american schools
- what s wrong with public education today