UNIVERSITY OF THE PUNJAB



BID DOCUMENTS

Selected bids are invited on Single Stage Two Envelopes (Technical & Financial) basis by University of the Punjab from manufacturers, authorized distributors, dealers, firms, registered with GST & NTN and PRA (if applicable), for the procurement of below mentioned “Laboratory Equipment (Gas Chromatograph, Inductively Coupled Plasma-Mass Spectrometer, XRF with Auto Sample Loading Position, Advance Diffractometer System etc.) for Centre for Coal Technology / Institute of Energy & Environmental Engineering”. Single Envelope Containing Two Separate Envelopes must be marked as “Financial Proposal” & “Technical Proposal”. {C&F and FOR Basis (In Pak Rupees Only)}

|Sr. |ITEMS DESCRIPTIONS |

|No. | |

|1 |Gas Chromatograph – Mass Spectrometer with Computer, Backup UPS, and UPS for Equipment |01 No. | | | |

| |Specifications (or equivalent) | | | | |

| |Oven | | | | |

| |Ballistic oven cool-down | | | | |

| |The ballistic cool-down time for the oven from 450°C to 50°C | | | | |

| |Oven and column characteristics | | | | |

| |Columns: Accepts multiple sizes of fused silica, packed or capillary columns coil. | | | | |

| |Oven temperature Range: 4°C above ambient to 450°C | | | | |

| |Settings: Temperature directly selectable in steps throughout the temperature range | | | | |

| |Average ambient sensitivity: Not more than 0.05°C change over a 5°C change in ambient | | | | |

| |temperature, within the 10 to 35°C allowable ranges | | | | |

| |High-temp protection: Firmware protection prevents safety hazard | | | | |

| |Oven temperature programming | | | | |

| |Temperature settings: Initial, final temperature, selectable with different increments | | | | |

| |Time settings: | | | | |

| |1 min increments for values 0 to 998 mins | | | | |

| |0.1 min increments for values 0 to 99.8 mins | | | | |

| |0.01 min increments for value 0 to 9.99 mins and infinite time (999 mins) | | | | |

| |Programming rate: Settable from 0.0°C/min to 160.0˚C/min in 0.1°C increments | | | | |

| |Pneumatic program rates: 0-100.0 psi/min, 0-100.0 mL/min, 0-200.0 cm/sec or ballistic | | | | |

| |Real-time graphical display of chromatogram and graphical display of temperature and | | | | |

| |pneumatic programs | | | | |

| |Injection countdown for manual injections | | | | |

| |Column pressure/flow/velocity calculator | | | | |

| |Meaningful error/alarm messages | | | | |

| |Log file | | | | |

| |Upgradable firmware | | | | |

| |Preventative-maintenance counter | | | | |

| |Password protection | | | | |

| |Status-summary screen | | | | |

| |Split/Splitless Injector | | | | |

| |Front capillary column injector | | | | |

| |Split ratio easily adjustable for a wide range of analytical conditions | | | | |

| |Charcoal trap in split vent prevents contamination of split valve and lab air | | | | |

| |Two choices of liner | | | | |

| |PPC pneumatics – four software configurable modes: programmed flow, programmed pressure, | | | | |

| |programmed velocity or constant flow. Vacuum compensation software selectable. | | | | |

| |PPC pneumatics include automatic control of split vent by split flow or split ratio | | | | |

| |Splitless liner | | | | |

| |Included with Split/Splitless Injector | | | | |

| |Low Bleed Column | | | | |

| |MS Low Bleed Column 30M-.25UM-.32MM | | | | |

| |Data System | | | | |

| |Includes the software CD, User’s Manual CD, and utilities CD. | | | | |

| |Compatible computer with 18.5 ' LCD, Laser Printer | | | | |

| |MASS SPECTROMETER | | | | |

| |Mass Spectrometer with EI | | | | |

| |Electron ionization with 75 L/sec turbomolecular pump | | | | |

| |Ion Source: Simplified Maintenance and Removal Technology | | | | |

| |Filament: designed for long life | | | | |

| |Mass range: 1.0-1200 u (amu) | | | | |

| |Detector: long-life high sensitivity electron multiplier detector | | | | |

| |Analyzer: Quadrupole with pre-filter | | | | |

| |Mass Stability: ±0.1 m/z over 48 hours | | | | |

| |EI Voltage: 10 – 100 eV | | | | |

| |Vacuum Pumps: 75 L/sec (nitrogen) air-cooled turbomolecular pump (59 L/sec helium pumping| | | | |

| |capacity) | | | | |

| |Vacuum Gauge: Single wide-range gauge | | | | |

| |Temperature Settings: Settable 50 ˚C – 350 ˚C | | | | |

| |Carrier Gas Flow Rate: 1.5 ml/min | | | | |

| |Mass Calibrant: FC-43 gas, triazine (high mass applications), user selectable | | | | |

| |MS Isolation: Channel Flow Technology | | | | |

| |MS Data Collection: Full Scan, Selected Ion Monitoring (SIM), Simultaneous Full-Scan | | | | |

| |Selected Ion Monitoring (SIFI) | | | | |

| |Scan Rate: Fully variable up to 12,500 amu/sec | | | | |

| |Acquisition Rate: 100 points/sec (SIM) | | | | |

| |Functions/Run: 32 functions or 32 ions per function | | | | |

| |Linear Dynamic Range: 106 dependent on acquisition rate | | | | |

| |Software: Compatible operating system | | | | |

| |Methods: Electronically transferable between GC/MS systems | | | | |

| |GC Acquisition: Full control and data processing of a single GC/MS with up to 2 GC | | | | |

| |detectors | | | | |

| |MS Tuning: Standard Tune environmental tuning optimized for EPA methods; Custom Tune user| | | | |

| |definable optimization | | | | |

| |Reporting: Included with specialized reports for environmental applications. | | | | |

| |Power: 230 VAC @ 50/60 | | | | |

| |Relative Humidity: 20 – 80 % non-condensing | | | | |

| |Also includes: | | | | |

| |• Wide range pressure gauge | | | | |

| |• 2.5 m3hr-1 backing pump | | | | |

| |• Cable, RS232 | | | | |

| |• Includes materials for installation. | | | | |

| |• Helium specific (multi-bed) quick connect filter glass indicating, multi bed gas | | | | |

| |filter kit with quick connects. | | | | |

| |• abrasive paper for ion source cleaning (6000 grit, 8000 grit:). | | | | |

| |ACCESSORIES | | | | |

| |Power Cable | | | | |

| |Tuning Compounds | | | | |

| |Ultra-micro Volume Syringe 5.0ul | | | | |

| |NUT KIT 1/8 STEEL PK/5 | | | | |

| |0.4mm Graphite / Vessel Ferrules Pack of 10 | | | | |

| |0.5mm Graphite / Vessel Ferrules Pack of 10 | | | | |

| |SPECIFIC TRIPLE GLASS TRAP CLICK-ON | | | | |

| |He Gas Regulator & Cylinder | | | | |

| |Libraries | | | | |

| |NIST Mass Spectral Library | | | | |

| |Includes the NIST CD, utilities CD, and the printed Install sheet. 190,825 spectra of | | | | |

| |163,198 compounds. Includes AMDIS deconvolution software. Required for chemical structure| | | | |

| |display, select environmental reports, and select custom reports. | | | | |

| |Main features: | | | | |

| |1. Electron Impact (EI) mass spectral library - 220,460 spectra of 192,108 unique | | | | |

| |compounds, with identifications and typically chemical structures. | | | | |

| |2. Gas Chromatography (GC) data library - 224,038 Kovats retention index values for | | | | |

| |21,847 compounds in the EI library, now on both polar and non-polar columns. Includes | | | | |

| |retention indices with GC column conditions and literature citations. | | | | |

| |3. NIST MS Search software - software for searching (identifying) compounds from their | | | | |

| |mass spectra and for browsing mass spectral libraries. Also includes MS interpretation | | | | |

| |programs for analyzing mass spectra on the basis of chemical structure, molecular | | | | |

| |formula, isotopic patterns, and more. | | | | |

| |4. AMDIS software - Automated Mass Spectral Deconvolution and Identification System for | | | | |

| |complex chromatograms. | | | | |

| |Note: Machine must include all accessory and other parts to install/make it operational. | | | | |

|2 |Inductively Coupled Plasma-Mass Spectrometer (ICP-MS) with Computer, Backup UPS, and UPS |01 No. | | | |

| |for Equipment | | | | |

| |Specifications (or Equivalent) | | | | |

| |Spectrometer | | | | |

| |Application software | | | | |

| |Installation Kit which includes: | | | | |

| |Installation Solution | | | | |

| |Gas Lines | | | | |

| |Cooling Lines | | | | |

| |Flexible Vent Tubing | | | | |

| |Waste Container | | | | |

| |Cyclonic sample introduction system | | | | |

| |Concentric nebulizer | | | | |

| |Torch | | | | |

| |Interface spares kit | | | | |

| |Built-in free-running Inductively Coupled Plasma source | | | | |

| |New cassette torch mount | | | | |

| |Quadruple-stage | | | | |

| |Metal quadrupole | | | | |

| |Thermally stabilized quadrupole power supply | | | | |

| |Peristaltic pump | | | | |

| |Computer-controlled gas pneumatics | | | | |

| |Detection limits | | | | |

| |Element ng/L (ppt) | | | | |

| |9Be < 1 | | | | |

| |59Co < 1 | | | | |

| |115In < 0.5 | | | | |

| |238U < 0.5 | | | | |

| |Sensitivity | | | | |

| |Element M cps/mg/L | | | | |

| |9Be > 3 | | | | |

| |24Mg > 20 | | | | |

| |115In > 50 | | | | |

| |238U > 40 | | | | |

| |Oxide and doubly charged species | | | | |

| |CeO+/Ce+ < 0.025 | | | | |

| |Ce++/Ce+ < 0.03 | | | | |

| |Background signal | | | | |

| |Mass 220 < 1 cps | | | | |

| |Short-term precision | | | | |

| |< 3% RSD | | | | |

| |Long-term stability | | | | |

| |< 4% RSD over 4 hours | | | | |

| |Isotope-ratio precision | | | | |

| |Defined for the isotope ratio of 107Ag/109Ag. Obtained using single-point peak hopping. | | | | |

| |< 0.08% RSD | | | | |

| |Mass calibration stability | | | | |

| |< 0.05 amu over 8 hours of continuous operation | | | | |

| |Carrier Gas Filter | | | | |

| |Necessary Tubings / Connectors included with main Unit | | | | |

| |Compatible computer and LCD/LED | | | | |

| |ICPMS Software Included with the main unit | | | | |

| |ICP-MS Checkout Solutions | | | | |

| |All required installation solutions should be included with main unit | | | | |

| |Other Standard Solution | | | | |

| |Kit-Solution Nexion Cell | | | | |

| |Solution Aft Single Element | | | | |

| |Solution Aft Multi-Element | | | | |

| |Multi-Element Calibration Standard | | | | |

| |Matrix per Volume: 5% HNO3 per 100 mL | | | | |

| |Heat exchanger, Air Cooled Recirculator  | | | | |

| |Refrigerated Chiller instead of Air Cooled Ciculator. Refrigerated Chiller is recommended| | | | |

| |where temperatrue goes above 30◦C. | | | | |

| |Coolant for Exchanger System (1 Litre) | | | | |

| |Autosampler for ICP | | | | |

| |Blower Vent Assembly | | | | |

| |Laser Printer | | | | |

| |Argon Humidifier With Bypass | | | | |

| |Filter Assy-W/R250 Regulator/Gauge | | | | |

| |Nebulizer | | | | |

| |Excellent mist and can handle high level of particulates in sample. | | | | |

| |Argon Gas Cylinder and Regulator | | | | |

| |Note: Machine must include all accessory and other parts to install/make it operational. | | | | |

|3 |XRF with Auto Sample Loading Position as per ASTM D 4294 with Computer, Backup UPS, and |01 No. | | | |

| |UPS for Equipment | | | | |

| |Specifications (or Equivalent) | | | | |

| |- Energy-dispersive X-ray spectrometer system for XRF analysis of the elements from | | | | |

| |Sodium to Uranium in solid, powder and liquid samples | | | | |

| |- Compact all-in-one spectrometer | | | | |

| |-Mains: 240 V, 50 Hz | | | | |

| |- X-ray tube: end-window, Pd anode | | | | |

| |- Primary beam filter changer: 9 positions | | | | |

| |- Detector: | | | | |

| |Peltier cooled solid state | | | | |

| |typical resolution ................
................