DIONEX ICS-3000 Ion Chromatography (IC) system

Switch back to ‘Home’ panel (Figure 9); Figure 9. Home panel. Click ‘Start up’ button, wait for 30 seconds or until pump pressure is stablized; Click ‘Start up’ again to ensure the status of EG component is ‘On’ (the eluent generate will not start if pump pressure is below 1000 psi) Make sure all components are working properly; ................
................