Quantitative Analysis of THC and Metabolites in Urine With ...
Quantitative Analysis of THC and Metabolites in Urine With a Simple, Fast, and Clean Oasis PRiME HLB ?Elution Plate
Dr. Xin Zhang, Dr. Jonathan P. Danaceau, and Dr. Erin E. Chambers Waters Corporation, Milford, MA, USA
APPLICATION BENEFITS Faster, simplified sample
preparation workflow Consistent recovery and minimal
matrix effects No evaporation or reconstitution necessary Linear, accurate, and precise results
for all analytes
WATERS SOLUTIONS Oasis PRiME HLB 96-well Elution Plate (p/n 186008052) 96-well Sample Collection Plate, 700 ?L, Round well (p/n 186005837) ACQUITY UPLC? BEH C18 Column, 130?, 1.7 m, 2.1 x 50 mm (p/n 186002350) ACQUITY UPLC I-Class System Xevo? TQ-S Mass Spectrometer
INTRODUCTION
Sample preparation is an important consideration for any bioanalytical LC-MS/MS method designed for forensic toxicology. Waters has developed a novel sample preparation sorbent, Oasis? PRiME, which is designed to have some key advantages over traditional SPE sorbents. These include the ability to eliminate sorbent preconditioning and equilibration, allowing a more rapid workflow compared to traditional SPE products, and the ability to remove more interferences, resulting in a cleaner extracts and reducing the risk of short column lifetimes or MS source fouling.
This application note details the extraction and UPLC-MS/MS analysis of -9-tetrahydrocannabinol (THC) and its metabolites, 11-hydroxy- -9-THC (THC-OH) and 11-nor-9-Carboxy--9-THC (THC-COOH) from urine using Oasis PRiME ?Elution Plates. -9-tetrahydrocannabinol (THC) is the main psychoactive element present in the plant Cannabis sativa. Quantitative analysis of these compounds in urine is an indicator of cannabis consumption, with high levels indicating recent and/or chronic use.
The use of Oasis PRiME resulted in consistent and highly reproducible recoveries of all compounds with minimal matrix effects. The Elution format allowed for the concentration of the sample on the SPE column, eliminating the need to evaporate and reconstitute the sample, minimizing the risk of analyte loss due to nonspecific binding and streamlining the laboratory workflow. This resulted in a method that was linear, accurate and precise for all analytes, with limits of quantification of 0.1 ng/mL for THC and its metabolites.
KEY WORDS -9-tetrahydrocannabinol (THC) and its metabolites, THC-COOH, THC-OH, urine, Oasis PRiME HLB, ?Elution plate, sample preparation, LC-MS
1
E X P E R IM E N TA L
Methods
All standards and stable isotope labelled internal standards were purchased from Cerilliant (Round Rock, TX, USA). Stock standards at 100 ?g/mL were prepared in 40% methanol (THC, THC-OH, and THC-COOH). A working internal standard solution of 1 ?g/mL THC-D3, THC-OH-D3 and THC-COOH-D3 was also prepared in 40% methanol. Individual calibrators and quality control standards were prepared daily in 40% methanol. 80 L of each working calibrator or QC standard was added to 1920 L of human urine to make calibration curves and QC samples.
-Glucoronidase from E. Coli K 12 was purchased from Roche Life Science (Indianapolis, IN)
Sample preparation Glucuronide hydrolysis: 40 ?L internal standards was added to 2 mL spiked human urine sample in a glass vial, then 2.4 mL 0.1 M potassium phosphate buffer (pH 6.8) containing 10 ?L -Glucoronidase was added. Vials were capped, vortex mixed, and incubated at 37 ?C water bath for 16 hours. After allowing samples to cool down to room temperature, 150 ?L of 10 M NaOH was added, vortex mixed and hydrolyzed in a dry heating block for 30 min at 70 ?C. Once the samples had cooled, 850 ?L glacial acetic acid was added to the samples and vortex mixed.
Solid-Phase Extraction with Oasis PRiME ?Elution Plate: 500 ?L pretreated sample (equivalent to 180 ?L urine) was directly applied to the Oasis PRiME Elution Plate. All wells of the SPE plate were then washed with 2 x 300 L aliquots of 25% methanol. The samples were then eluted with 2 x 25 L aliquots of 60:40 ACN:IPA and diluted with 50 L of water. 5 L was injected onto the UPLC-MS/MS system. The SPE extraction procedure is summarized in Figure 1.
Analyte recovery was calculated according to the following equation:
( ) %Recovery = Area A x 100% Area B
Where A equals the peak area of an extracted sample and B equals the peak area of an extracted blank matrix sample in which the compounds were added post-extraction.
Matrix effects were calculated according to the following equation:
(( ) ) Matrix Effects = Peak area in the presence of matrix ?1 x 100% Peak area in the absence of matrix
The peak area in the presence of matrix refers to the peak area of an extracted matrix sample in which the compounds were added post-extraction. The peak area in the absence of matrix refers to analytes in a neat solvent solution.
LC conditions LC system:
ACQUITY I-Class UPLC System
Column:
ACQUITY UPLC BEH C18 Column, 130?, 1.7 m, 2.1 x 50 mm
Column temp.:
40 ?C
Sample temp.:
10 ?C
Mobile phase A (MPA): Water with 0.1% formic acid
Mobile phase B (MPB): ACN with 0.1% formic acid
Strong wash solvent: 70:30 ACN:Water with 2% formic acid
Weak wash solvent: 10% ACN
Injection vol.:
5 ?L
The gradient ramp is shown in Table 1.
Time
Flow
(min) (mL/min) %A
%B
0
0.6
50
50
1.0
0.6
50
50
3.0
0.6
5
95
3.5
0.6
5
95
3.6
0.6
50
50
4.0
0.6
50
50
Table 1. Mobile phase gradient. The compositions of MPA and MPB are listed in the Methods section.
Mass spectrometry
MS system:
Xevo TQ-S Mass Spectrometer
Ionization mode:
ESI Positive
Capillary voltage: 2.0 kV
Cone voltage:
Optimized for each analyte
Desolvation gas:
1000 L/hr
Cone gas:
150 L/hr
Desolvation temp.: 500 ?C
Source temp.:
150 ?C
Data were acquired and analyzed using MassLynx? Software (v4.1). Quantification was performed using TargetLynxTM.
Quantitative Analysis of THC and Metabolites in Urine With a Simple, Fast and Clean Oasis PRiME HLB ?Elution Plate
2
RESULTS AND DISCUSSION
Chromatography Figure 2 shows chromatography of the three cannabinoids from an extracted calibrator at 2 ng/mL. All compounds eluted within 3 minutes with all peak widths were under 3 seconds at 5% of baseline. All peaks were symmetrical with symmetries between 0.95?1.15.
Table 2 lists the retention time and individualized MS parameters of the cannabinoids and their stable isotope labelled internal standards, including MRM transitions, cone voltage, and collision energy. Two MRM transitions were used for each compound, a primary (listed first) and a confirmatory transition (listed second).
Load Prepared Urine Sample
Wash 2 x 300 ?L 25% MeOH
Elute 2 x 25 ?L (60:40 ACN:IPA)
Figure 1. Oasis PRiME extraction methodology for urine THCs. With no conditioning and equilibration, sample extraction is simplified to just three steps.
%
%
%
100
0 1.00 100
0 1.00 100
0 1.00
1.84
THC-OH
1.50
2.00
2.50
1.92
THC-COOH
3.00
1.50
2.00
2.50
3.00
Figure 2. Chromatography
THC
2.72
of THC-OH, THC-COOH and THC from an extracted urine
sample on the ACQUITY
UPLC BEH C18 column,
1.8 ?m; 2.1 x 50 mm. The
1.50
2.00
2.50
Time 3.00
concentrations are 4 ng/mL
for all compounds.
Analyte
RT (min)
THC-OH
1.84
THC-OH-d3
1.84
THC-COOH
1.92
THC_COOH-d3 1.92
THC
2.72
THC-d3
2.72
MRM transitions (m/z)
331.3>313.1 331.3>193.1 334.3>316.1 345.3>327.3 345.3>299.3 348.3>330.3 315.1>193.2 315.1>135.1 318.1>196.2
Cone voltage
(V)
40 40
40
50 50
50
40 40
40
Table 2. Mass spectral parameters for all analytes and internal standards.
Collision energy
(eV)
18 30
18
20 25
20
25 25
25
Quantitative Analysis of THC and Metabolites in Urine With a Simple, Fast and Clean Oasis PRiME HLB ?Elution Plate
3
Recovery and matrix effects
Extraction recoveries were very consistent. As Figure 3 shows, recovery for THC-OH and THC-COOH was around 90% and THC was 60% with all %RSDs under 7.5%, demonstrating the reproducibility of Oasis PRiME. Matrix effects were minimal, at less than 15% for all compounds. Once again, the low standard deviations (7.5% or less) and high recoveries for THC-OH and THC-COOH demonstrate the consistency of extraction and cleanup seen with Oasis PRiME HLB. All recovery and matrix effect data are summarized in Table 3. Oasis PRiME HLB also provided better recovery, variability and matrix effects than LLE, with a more simplified procedure.1
120 100
80 60 40 20
0 -20
THC-OH
Recovery Matrix Effect
THC-COOH
THC
Figure 3. Recovery and matrix effects of THC-OH, THC-COOH, and THC after extraction using the Oasis PRiME ?Elution plate. %RSDs for extraction recovery were less than 5% for all compounds. Matrix effects were all within 20%.
Quantitative results
Calibration and quality control samples were prepared as previously described in the materials and method section. Calibration ranges were from 0.1?100.0 ng/mL for THC-COOH and THC-OH and 0.2?100.0 ng/mL for THC. Quality control samples were prepared at low, medium, and high concentrations as appropriate for the calibration ranges.
All compounds had linear responses over the entire calibration range with R2 values of 0.99 or greater with 1/x weighting. Table 4 summarizes the data from the calibration curves. Lower limits of quantification (LLOQ) were 0.1 ng/mL for THC-COOH and THC-OH and 0.2 ng/mL for THC. In each case, all FDA recommendations for accuracy, precision, linearity and analytical sensitivity were met for validated methods.2
Quality control samples were accurate and precise. All results were within 15% of expected values and %RSDs were under 2% (N=6). This data can be seen in Table 5. The excellent accuracy and precision demonstrate the consistency and robustness of this sorbent.
THC-OH THC-COOH THC
Mean 91.9 91.5 60.6
Recovery
S.D. 4.5 4.5 4.5
%RSD 4.9 4.9 7.4
Matrix effects
Mean
S.D.
-0.4
3.0
-10.4
3.7
-12.9
4.5
Table 3. Recovery and Matrix effects for THC and its metabolites (N=4 for all tests).
THC-OH THC-COOH THC
R2 0.997 0.998 0.998
Mean % dev. 2.8 2.0 1.2
Table 4. Calibration Curve Summary for THC and its metabolites with 1/x fit weighting.
Quantitative Analysis of THC and Metabolites in Urine With a Simple, Fast and Clean Oasis PRiME HLB ?Elution Plate
4
N=6
QC level (ng/mL)
0.75 7.5 75 Mean
Mean (ng/mL)
0.66 6.70 73.4
THC-OH
%Acc.
88.6 89.3 97.9 91.7
%RSD
1.7 1.3 1.8
Accuracy and precision
THC-COOH
Mean (ng/mL)
%Acc. %RSD
0.76
100.8 1.4
7.37
98.3 1.3
73.6
98.2 0.7
99.7
Mean (ng/mL)
0.72 7.15 75.5
Table 5. Quality control results from extracted urine samples. (N=6 for each compound at all three levels).
THC
%Acc.
96.3 95.3 100.7 97.4
%RSD
0.4 1.2 0.8
CONCLUSIONS Calibration and quality control samples were prepared as previously described in the materials and method section. Calibration ranges were from 0.1?100.0 ng/mL for THC-COOH and THC-OH and 0.2?100.0 ng/mL for THC. Quality control samples were prepared at low, medium, and high concentrations as appropriate for the calibration ranges.
All compounds had linear responses over the entire calibration range with R2 values of 0.99 or greater with 1/x weighting. Table 4 summarizes the data from the calibration curves. Lower limits of quantification (LLOQ) were 0.1 ng/mL for THC-COOH and THC-OH and 0.2 ng/mL for THC. In each case, all FDA recommendations for accuracy, precision, linearity and analytical sensitivity were met for validated methods.2
Quality control samples were accurate and precise. All results were within 15% of expected values and %RSDs were under 2% (N=6). This data can be seen in Table 5. The excellent accuracy and precision demonstrate the consistency and robustness of this sorbent.
References
1. Lee, R., Traynor, A., LeCount, J., Wood, M., Quantitative analysis of 11-nor-carboxy 9-THC in urine using UPLC-MS/MS. Waters Application Note 720004808EN (2012).
2. Bansal, S., DeStefano, A., Key elements of bioanalytical method validation for small molecules. The AAPS Journal 9(1), E109-E114 (2007).
Waters, T he Science of W hat's Possible, Oasis, ACQUITY, ACQUITY UPLC, MassLynx, and Xevo are registered trademarks of Waters Corporation. TargetLynx is a trademarks of Waters Corporation. All other trademarks are the property of their respective owners.
?2015 Waters Corporation. Produced in the U.S.A. December 2015 720005556EN AG-PDF
Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990
................
................
In order to avoid copyright disputes, this page is only a partial summary.
To fulfill the demand for quickly locating and searching documents.
It is intelligent file search solution for home and business.
Related download
- quantitative analysis of thc and metabolites in urine with
- extraction of thc and metabolites from urine and plasma
- e t drug court u practitioner i national drug court
- detection times of drugs of abuse in blood urine and
- quantitative analysis of thc and main metabolites in whole
- urine drug testing for chronic pain management
- urine drug testing
- thc metabolite elisa kit
- thc cooh quantification in urine usingdilute and shoot lc
Related searches
- list of colleges and universities in florida
- list of colleges and universities in illinois
- thc metabolites in urine test
- metabolites in urine testing
- number of colleges and universities in usa
- meth metabolites in urine testing
- alcohol metabolites in urine testing
- definition of quality and safety in nursing
- ministry of labour and employment in jamaica
- list of colleges and universities in usa
- list of cities and counties in texas
- blood in urine with catheter